4-17 Calibrating the D/A in 10Hz

100Hz range of SW MODE

n Set voltage of POWER SUPPLY to 60V and current of it to 1A.
o Connect the positive output terminal of POWER SUPPLY to a 0.1Ω 5W concrete resistance first, then

connect the resistance to the input terminal of the electronic load. Hook up a DMM in 200V range with
the input terminal of the load. Connect the input terminal of COUNTER to both ends of the concrete
resistance.

p Test the input terminal of the load by the 200V range of a DMM. Press SHIFT and C.V. MODE, then key

in 3, 8, 1,6, and

↵ . Then adjust the POWER SUPPLY until the DMM display reads 60.00V.

q When MEMORY indicates “01” and A indicates “CL16”, key in the frequency reading (7 digits including

decimal) of the COUNTER (in unit of “Hz”) and press

↵ .

r When MEMORY indicates “02”, key in the frequency reading (7 digits including decimal) of the

COUNTER (in unit of “Hz”) and press

↵ .

s When MEMORY indicates “03”, key in the frequency reading (7 digits including decimal) of the

COUNTER (in unit of “Hz”) and press

↵ .

t When MEMORY indicates “04”, key in the frequency reading (7 digits including decimal) of the

COUNTER (in unit of “Hz”) and press

↵ .

u When MEMORY indicates “05”, key in the frequency reading (7 digits including decimal) of the

COUNTER (in unit of “Hz”) and press

↵ .

v When MEMORY indicates “06”, key in the frequency reading (7 digits including decimal) of the

COUNTER (in unit of “Hz”) and press

↵.

Note: Applied instruments -- POWER SUPPLY (60V, 6A); DMM (5 1/2 digits);

COUNTER (frequency counter)

Applied situation --The readings of FREQ. and DUTY are not reliable.