beautypg.com

GW Instek PEL-300 User Manual

Page 33

background image

30

4-16 Calibrating the100Hz

1kHz range of SW MODE

n Set voltage of POWER SUPPLY to 60V and current of it to 1A.
o Connect the positive output terminal of POWER SUPPLY to a 0.1Ω 5W concrete resistance first, then

connect the resistance to the input terminal of electronic load. Hook up a DMM in 200V range with the
input terminal of the load. Connect the input terminal of COUNTER to both ends of the concrete
resistance.

p Test the input terminal of the load by the 200V range of a DMM. Press SHIFT and C.V. MODE, then key

in 3, 8, 1,5, and

↵ . Then adjust the POWER SUPPLY until the DMM display reads 60.00V.

q When MEMORY indicates “01” and A indicates “CL15”, key in the frequency reading (7 digits including

decimal) of the COUNTER (in unit of “Hz”) and press

↵ .

r When MEMORY indicates “02”, key in the frequency reading (7 digits including decimal) of the

COUNTER (in unit of “Hz”) and press

↵ .

s When MEMORY indicates “03”, key in the frequency reading (7 digits including decimal) of the

COUNTER (in unit of “Hz”) and press

↵ .

t When MEMORY indicates “04”, key in the frequency reading (7 digits including decimal) of the

COUNTER (in unit of “Hz”) and press

↵ .

u When MEMORY indicates “05”, key in the frequency reading (7 digits including decimal) of the

COUNTER (in unit of “Hz”) and press

↵ .

v When MEMORY indicates “06”, key in the frequency reading (7 digits including decimal) of the

COUNTER (in unit of “Hz”) and press

↵ .

Note: Applied instruments -- POWER SUPPLY(60V, 6A); DMM(5 1/2 digits); COUNTER(frequency counter)

Applied situation -- The readings of FREQ. and DUTY are not reliable.