4-2 Adjusting the reference voltage in 10

Ω∼1k range of C.R. MODE

n Set voltage of POWER SUPPLY to 60V and current of it to 1A.
o Connect the output terminal of POWER SUPPLY to the input terminal of electronic load.
p Press SHIFT and C.V. MODE, then key in 3, 8, 0, 2, and ↵.
q Test the input terminal of load by the 200V range of a DMM. Adjust the POWER SUPPLY until the DMM

display reads 60.00V. Use the 20V range of another DMM to test TP1 of a PCB.

r When MEMORY indicates “01” and A indicates “CL02”, key in the voltage reading (no less than 6 digits

including decimal) of TP1 (in unit of “V”) and press

↵.

s When MEMORY indicates “02”, key in the voltage reading (no less than 7 digits including decimal) of TP1

(in unit of “V”) and press

↵ .

t When MEMORY indicates “03”, key in the voltage reading (no less than 7 digits including decimal) of TP1

(in unit of “V”) and press

↵ .

Note: Applied instruments -- POWER SUPPLY (60V, 6A); DMM (5 1/2 digits)

Applied situation -- The C.R. MODE in 10

Ω∼1kΩ range is not reliable.

4-3 Calibrating the D/A in 10

range of C.R. MODE

n Set voltage of POWER SUPPLY to 17V and current of it to 20A.
o Connect the output terminal of POWER SUPPLY to a DMM in 20A range in series. Hook up the DMM

with the input terminal of the electronic load.

p Test the input terminal of the load by the 20V range of a DMM. Press SHIFT & C.V. MODE, key in 3,8,0,3,

and

↵. Then adjust the POWER SUPPLY until the DMM display reads 17.00V.