GW Instek PEL-300 User Manual

Page 16

background image

13

(7). Setting of dynamic mode:

n Connect the instrument according to Graph 3-3.
o Turn the power switch to “ON” position, then press C.C. MODE button (the light of C.C. MODE is on).
p Press SW ON/OFF button; the light of SW is on which indicates the operation is proceeded in dynamic

mode.

Note: be sure to complete the settings of FREQ. and DUTY before setting the dynamic mode.

Example:

To set operation in dynamic mode

STEP 1. Press SW ON/OFF button (the light of SW is on)
STEP 2. Press LOAD ON/OFF button

(the light of LOAD is on which indicates the operation is in SW mode)

(8). Setting of LOAD ON/OFF:

n Turn the power switch to “ON” position, then press C.C. MODE, C.V. MODE or C.R. MODE button.
o Press LOAD ON/OFF button; the light of LOAD is on which indicates the load has been turned on.
p Press LOAD ON/OFF button again; the light of LOAD is out which indicates the load has been turned

off.

Note: Make sure the preset input power is not too low that may cause an error message.