(3). Constant Resistance Mode:

n Connect the instrument according to Graph 3-3.
o Turn the power switch to “ON” position, then press C.R. MODE button (the light of C.R. MODE is on).
p Press LOAD ON/OFF button (the light of LOAD is on), the displays will indicate the value of external

voltage and current.

Example:

To set input resistance at 500

Ω in constant resistance mode

STEP 1.

Press C.R. MODE

STEP 2.

Press V.I.R. SET

STEP 3.

Press 500 (unit is specified as “

Ω” )

STEP 4.

Press

(4). Setting of maximum input power:

n Select C.V. MODE, C.C. MODE or C.R. MODE.
o When the light of the selected mode is on, press W SET button. Input desired W value after the

display indicating the original W value.

p Press LOAD ON/OFF button.
Example:

To set maximum input power at 300W

STEP 1.

Press W SET

STEP 2.

Press 300 (unit is specified as “W” )

STEP 3.

Press