4-14 Calibrating the D/A in 100

range of C.R. MODE

n Set voltage of POWER SUPPLY to 40V and current of it to 6.2A.
o Connect the output terminal of POWER SUPPLY to a DMM in 20A range in series. Then hook up the

DMM with the input terminal of the electronic load.

p Test the input terminal of the load by the 200V range of a DMM. Press SHIFT and C.V. MODE, then key

in 3, 8, 1, 3, and

↵ . Adjust the POWER SUPPLY until the DMM display reads 40.

q When MEMORY indicates “01” and A indicates “CL13”, key in the current reading (no less than 6 digits

including decimal) of the DMM (in unit of “A”) and press

↵.

r When MEMORY indicates “02”, key in the current reading (no less than 6 digits including decimal) of the

DMM (in unit of “A”) and press

↵ .

s When MEMORY indicates “03”, key in the current reading (no less than 6 digits including decimal) of the

DMM (in unit of “A”) and press

↵ .

t When MEMORY indicates “04”, key in the current reading (no less than 6 digits including decimal) of the

DMM (in unit of “A”) and press

↵ .

Note: Applied instruments -- POWER SUPPLY (60V, 6A); DMM (5 1/2 digits)

Applied situation -- The readings of C.R. MODE are not reliable.