GOS-6051/6050/6031/6030 OSCILLOSCOPE

USER MANUAL

⎯ 34 ⎯

4. Set both CH1 and CH2 COUPLING to GND.

5. Use the CH1 and CH2 POSITION controls to align both traces on the center

graticule.

6. Open the CH2 INV by pressing and holding the pushbutton.
7. Set to the ADD mode by pressing the ADD pushbutton briefly.
8. Set both CH1 and CH2 COUPLING to DC.

9. The figure 5-4 below shows a satisfactory display. The display will show a

flat trace located on the center gracticule that confirms the channel balance
and ADD offset function.

Figure 5-4 ADD mode

10. Turn off the ADD mode by pressing the ADD pushbutton briefly.

11. Turn off the CH2 INV by pressing and holding the pushbutton.

GOS-6051/6050/6031/6030 OSCILLOSCOPE

USER MANUAL

⎯ 35 ⎯

5-5.Basic Operation

Displaying CH1 or CH2
To display the signal from a signal channel, pressing briefly the CH1 or CH2
pushbutton to set the oscilloscope to channel 1 or channel 2.
Displaying CH1 and CH2
To display both signals at the same time, proceed the following steps:

1.Set the CH1 and CH2 on. The figure 5-5 below shows two synchronous

waveforms in the both modes.

2.Adjust the CH1 or CH2 POSITION control to position the two waveforms.
3.Set the ALT/CHOP button to CHOP mode if the waveforms are flickering.

Figure 5-5 Both typical waveforms