beautypg.com

GW Instek GOS-6000 Series User Manual

Page 19

background image

GOS-6051/6050/6031/6030 OSCILLOSCOPE

USER MANUAL

⎯ 32 ⎯

shown in figure 5-2. If either probe needs to be adjusted, proceed the step 8.
If either probe does not need to be adjusted, proceed the “Function Check”.

Figure 5-2 Typical Compensation Waveform

8.Adjust the probe by using a small insulated screwdriver. Slowly rotate the

adjustment control until the probe is properly compensated.

GOS-6051/6050/6031/6030 OSCILLOSCOPE

USER MANUAL

⎯ 33 ⎯

5-4.Function Check

When you start to check the operation of your oscilloscope, proceed the
following instruction:
1. Install the ×10 probes onto CH1 and CH2 inputs.
2. Connect the probe tips to the CAL test point of the oscilloscope.
3. Set the oscilloscope controls to display both channels:

VERTICAL:

VOLTS/DIV

0.2V

COUPLING

DC

ALT/CHOP

CHOP

HORIZONTAL:

TIME/DIV

0.5ms

TRIGGER:

MODE

ATO

SOURCE

VERT

COUPLING

AC

SLOPE

The figure 5-3 below illustrates a satisfactory display. The waveform should
be approximately 0.5Vp-p at a frequency of 1kHz that confirms the vertical
and horizontal deflection function of the oscilloscope.

Figure 5-3