Aquametrix AM-TBR Turbidimeter User Manual User Manual
Page 20

15
12. Insert the requested 0.02 NTU standard. Index the standard to the lowest value on the
upper display.
13. Press the ↵ button to accept the calibration.
14. The lower display will count down the progress of the calibration step.
15. The instrument will return to AUTO mode at the end the calibration.
Note: During calibration, the fan inside the instrument is turned off to extend the life of the
desiccant. The fan will be turned on during calibration countdowns and after returning to the
AUTO mode or after five minutes, which ever comes first. It is recommended that the
measurement chamber be kept covered during the calibration period and that the flow through
cuvette be replaced immediately after the calibration to prevent premature saturation of the
desiccant.
5.3. Calibration Error
If the screen shown below, is displayed after calibration, the internal diagnostics have
determined that the calibration standards were either bad or that they were inserted in the wrong
order. Either check the standards and recalibrate or restore the factory calibration see 6.2. The
instrument cannot be used without performing one of these operations.
To recalibrate press the MODE key and start the calibration sequence again. To restore the
factory calibration, push and hold the button. Now push and release the ↵ then release
the button.