Aquametrix AM-TBR Turbidimeter User Manual User Manual

Page 19

background image

14

The arrow beside CAL will be illuminated on the display. The lower display shows
alternating 1000 (the value of the standard that is requested) and ↵. The upper display
shows the real-time reading to allow the standard to be indexed. Refer to section 6.1 for
information on indexing cuvettes.

2. Remove the flow through unit.

3. Insert the requested 1000 NTU standard. Index the standard to the lowest value on the

upper display.

4. Press the ↵ button to accept the calibration.
5. The lower display will count down the progress of the calibration step.

6. The lower display will now change to show alternating 10 and ↵, requesting the 10.0 NTU

standard.










7. If the alternating 10 and ↵ is not displayed, push the or

until this display is shown.

8. Insert the requested 10.0 NTU standard. Index the standard to the lowest value on the

upper display.

9. Press the

button to accept the calibration.

10. The lower display will count down the progress of the calibration step.

11. The lower display will now change to show 02 and

, requesting the 0.02 NTU standard.