3 checking/changing the text line terminator, 5 setting the ethernet communication parameters, Section 2.5 – LumaSense Technologies 3434i SF6 Leak Detector User Manual
Page 19

Chapter 2
______________________________________________________________________
_____________________________________________________________________________
BE6028-14
3434i SF6 Leak Detector
LumaSense Technologies A/S
Page 19 of 191
2.4.3
Checking/Changing the Text line Terminator.
The Text line Terminator must be set using the push-buttons on the
front of the monitor.
Press SET-UP S2 ▼ ▼ ▼S3
Press SET-UP to exit the set-up mode
2.4.4
Checking/Changing the Data Log and Error Log
Options.
The Data Log and Error Log must be set using the push-buttons on
the front of the monitor.
Press SET-UP S3 S1 S1 S3 ▼S1 S1
Press SET-UP to exit the set-up mode
2.5 Setting the Ethernet Communication Parameters
The LumaSoft Gas 7810 application with Ethernet communication is
able to communicate with the 3434i gas monitor through a local
Ethernet network.
The homepage of the 3434i gas monitor can also be displayed in a
web browser, when the 3434i gas monitor is connected to a local
Ethernet network.
2.5.1
Checking/Changing the Ethernet Communication
Parameters
The communication parameters for the Ethernet interface can be set
by using the push-buttons on the front of the monitor.
1. Press SET-UP S3 S1 S3 S2. The screen display now shows the
following text.
2. Press S1 to enter the Ethernet setup. During this setup you can
also use the ▲ and ▼ keys to go back and forth between the
SELECT ADDRESSABLE SET-UP BRANCH
Ethernet IEEE 488