1 producing a supply of clean, wet air – LumaSense Technologies 3434i SF6 Leak Detector User Manual
Page 146

Chapter 12
______________________________________________________________________
_____________________________________________________________________________
BE6028-14
3434i SF6 Leak Detector
LumaSense Technologies A/S
Page 146 of 191
tube. This condition must be avoided because atmospheric air will
be drawn into the Teflon tube system via the flow meter and
cause dilution of the calibration gas. This will affect the accuracy
of your calibration.
·
It functions as an escape valve. When the pressure of the gas in
the tubing becomes greater than atmospheric pressure, gas flows
out to the atmosphere via the flow meter. This ensures that the
gas entering the analysis cell is as close to atmospheric pressure
as possible, and this prevents damage to the very sensitive
microphones in the analysis cell.
WARNING!: The analysis cell of the monitor is equipped with
sensitive microphones, and therefore NO direct connection between
the Monitor’s air-inlet and the pressure-valve on a gas cylinder is
allowed. Under NO circumstances should the pressure of the air in
the analysis cell be allowed to exceed 0.1 bar above the ambient
pressure.
1. Attach the Teflon tubing to the Air Inlet Filter at the back of the
monitor. This is described in detail in
2.
a. Connect the free-end of the length of tubing attached to the
air-inlet to one of the branches of the “Y”-piece (see
b. Cut off another 1m length of Teflon tubing and connect one
of its ends to attachment point no. 2 of the “Y”-piece (see
) and its other end to a flow-meter.
3. Connect a 1m length of tubing to attachment point no. 3 of the
“Y”-piece (see
12.4.1 Producing a Supply of Clean, Wet Air
When humidity-interference calibrations are performed, clean, wet
air is required. This can be produced in the following manner.
We suggest that you bubble zero-gas (e.g. pure nitrogen) through a
thermostatically-controlled water-bath to produce a supply of clean,
wet air to the Monitor during humidity-interference calibration of the
filters (see
). Notice that the thermostatically-controlled
water bath is linked up to an extra “empty” flask. This is to prevent a
situation where the level of water in the controlled water-bath rises
and covers the outlet tube “A” and draws water, via the “Y”-piece,
directly into the Monitor’s analysis cell. Water will seriously damage
the cell. To avoid such a situation, we suggest that the wet air from
the water-bath outlet tube is fed into an extra empty flask before
being fed to the air-inlet of the Monitor. Note the length of the