Appendix 1 Specifications of PR300
Environmental Conditions
Normal operating conditions
Warm-up time
At least 30 minutes
Ambient temperature
0 to 50
°C (reference temperature: 23Ϯ2°C)
Temperature change
10
°C/h or less
Ambient humidity
20 to 90% RH (no condensation)
Magnetic field
400A/m or less
Continuous vibration
10 to 60 Hz, 0.035 mm, 75 minutes; 60 to 150 Hz, 4.9 m/s
2
, 75 minutes
Short-time vibration
14.7 m/s
2
for 15 seconds or less
Shock
98 m/s
2
or less (for shock time of 11 ms)
Installation
Indoor installation only
Mounting position
Vertical surface mounting only
Installation altitude
2000 m or less
Effect on operating conditions
Effects of ambient temperature
Analog output:
Ϯ0.05% of F.S./°C or less
Effects on power supply voltage variations
Analog output:
Ϯ0.05% of F.S./°C or less
Transport and storage conditions
Temperature
-20 to 70
°C
Humidity
5 to 95% RH (no condensation)
Shock and dropping of package
90 cm (provided that an external packing box is used)
Mounting and Shape
Materials
Casing
Polycarbonate resin (PC), UL94 V-0
Terminal block
Polybutylene terephthalate (PBT), UL94 V-0
Terminal cover
Polyamide resin (PA6), UL94 V-2
Mounting method
Panel mounting
Connection Method
M3 screws for terminal connections (analog output, pulse output, demand alarm output, digital
input, and RS-485 communication)
M4 screws for terminal connections (voltage/current input and power supply)
RJ45 connection (Ethernet communication, when the Ethernet communication function is specified)
External Dimensions (including a terminal cover)
110
ϫ 110 ϫ 128 mm (H ϫ W ϫ D)
96
ϫ 96 ϫ 126 mm (H ϫ W ϫ D)
Weight
Approx. 600 g (when accessories such as mounting bracket are attached)