<5. Parameter Summary>

5-11

IM 01C25R02-01E

Function

Label

Item

Contents

Default value Handling

*

1

Status

Status group 1

Device status information for

hardware

Display the hardware error

R

Status group 2

Device status information for

hardware

Display the hardware error

R

Status group 3

Device status information for

process

Display the process error

(Out of specification)

R

Status group 4

Device status information for

process

Display the process error

(Out of setting range)

R

Status group 5

Device status information for

process

Display the process error

(Alarm)

R

Status group 6

Device status information for data

Display the setting error

R

Status group 7

Device status information for data

Display the setting error

R

Status group 8

Device status information for data

Display the setting error and

process error (Alarm)

R

Status group 9

Device status information for data

Display the diagnostic alarm

R

Status group 10

Device status information for data

Display the diagnostic alarm

R

Status group 11

HART 7

Device status information for data

Display the simulation mode

R

Ext dev status

HART 7

Extended Device Status

R

Tag

Tag

Tag number

16 alphanumerics

As specified

W

Long tag

HART 7

Long tag

Max 32 alphanumerics

As specified

W

Test key

Test key

Special maintenance parameter

M

Total Flow

Cvt Val

Convert value for total flow base

unit

W

Freq at 100%

Digtal otput fequency at 100%

W

Pulse rate

Pulse rate

W

Modify Unit

Modify flow base unit

M

Set Base Unit

Set total flow base unit

M

Total Flow Mode

Total flow mode

Reset, Start, Start Cyclic or

Stop

W

Total Flow Unit

Current total flow unit

W

Units

Density Unit

Density unit

W

ET Unit

External temperature unit

W

Flow Unit

Flow unit

W

Pres Unit

Differential pressure unit

W

PV Unit

PV unit

W

SP Unit

Static pressure unit

W

SV Unit

SV unit

W

Total Flow Unit

Total flow unit

W

TV Unit

TV unit

W

4V Unit

HART 5

4V unit

W

QV Unit

HART 7

QV unit

W

Write

protection

menu

Enable wrt 10min

Write protection release

8 alphanumerics

M

New password

User set password for write

protection

8 alphanumerics

M

Write protect

Write protection indicator

Yes or No

No

R

*1: Handling: R-Read only, W-Read & Write, M-Method, G-Applicable for option code DG6

*2: Max three Burst Messages (Burst Message 1, Burst Message 2, Burst Message 3)