IM 01C25R02-01E

• Procedure to call up the display for differential

pressure

DD (HART 5/7)

DTM (HART 7)

[Root Menu] → Diag/Service →

Calibration → Pres Sensor trim →

Clear P trim → Execute

DTM (HART 5) Calibration → Clear Pressure Sensor

trim → Execute

• Procedure to call up the display for static

pressure

DD (HART 5/7)

DTM (HART 7)

[Root Menu] → Diag/Service →

Calibration → SP Sensor trim →

Clear SP trim → Execute

DTM (HART 5) Calibration → Clear Static Pressure

Sensor trim → Execute

• Procedure to call up the display for external

temperature

DD (HART 5/7)

DTM (HART 7)

[Root Menu] → Diag/Service →

Calibration → ET Sensor trim →

Clear ET trim → Execute

DTM (HART 5) Calibration → Clear External Temp

Sensor trim → Execute

3.3.7 Trim Analog Output

Fine current output adjustment is carried out with

D/A trim or Scaled D/A trim.

(1) D/A Trim
D/A trim
is to be carried out if the calibration digital

ammeter does not exactly read 4.000 mA and

20.000 mA with an output signal of 0% and 100%.

• Procedure to call up the

D/A trim display

DD (HART 5/7)

DTM (HART 7)

[Root Menu] → Diag/Service →

Calibration

→ Analog output trim → D/A trim

DTM (HART 5) Calibration → D/A trim

(2) Scaled D/A Trim
Scaled D/A trim
is to be carried out if the output is

adjusted using a voltmeter or a meter whose scale

is 0 to 100%.

• Procedure to call up the

Scaled D/A trim

display

DD (HART 5/7)

DTM (HART 7)

[Root Menu] → Diag/Service →

Calibration

→ Analog output trim →Scaled D/A

trim

DTM (HART 5) Calibration → Scaled D/A trim

Adjustment using a volt meter. (4mA → 1V, 20mA

→ 5V)

1) Select “Change”.
2) Enter the value read on the voltmeter when the out-

put signal is 4mA.

In this case, enter the value of the voltage across a
250Ω resistor (1V).

3) Enter the value read on the meter when the output

signal is 20mA (5V).

4) Select “Proceed”.
5) Connect the voltmeter.
6) Output the 0% output signal and read the output

value.

7) Enter the reading of the voltmeter to the configuration

tool. (The output of the transmitter changes).

8) Confirm the voltmeter reading is 1.000.
9) If the reading on the voltmeter is 1.000, select “Yes”.

If the reading is not 1.000, select “No” and repeat
steps 6 and 7 until the voltmeter reads 1.000V.

10) Output the 100% output signal and read the output

value.

11) Enter the reading of the voltmeter.
12) Confirm the voltmeter reading is 5.000.
13) If the reading of the voltmeter is 5.000, select “Yes”.

If the reading on the voltmeter is not 5.000, select
“No” and repeat steps 10 and 11 until the voltmeter
reads 5.000V.

3.3.8 External Switch Mode

Follow the procedure below to enable or inhibit zero

point adjustment by means of the zero-adjustment

screw on the transmitter.
This is set to “Enabled” when the instrument is

shipped.
To change the mode, follow the procedure below.

• Procedure to call up the display

DD (HART 5/7)

DTM (HART 7)

[Root Menu] → Detailed setup →

Device information → Field device info

→ Ext SW

DTM (HART 5) Configuration → Device information1

→ Ext SW

Enabled

Enable the external zero point

adjustment

Disabled

Disable the external zero point

adjustment