4.2.3 Heat Trace Monitoring

The EJX multivariable transmitter with Heat

trace monitoring function calculates the flange

temperature by using the two temperature sensors

built in the EJX multivariable transmitter.
An analog alert is generated if the temperature

reached to the preset level.
The flange temperature is based on the following

parameters and calculation formula.

[Parameters]

Parameter name

Explanation

Cap temp (CT)

Measured capsule temperature

value

Amp temp (AT)

Measured amplifier temperature

value

Flg temp (FT)

Flange temperature value

(Calculated value)

Flg temp Coef (Cf)

Coefficient to calculate flange

temperature

Flg temp Hi Alert Val Threshold to generate FT high

alarm

Flg temp Lo Alert Val Threshold to generate FT low

alarm

[Calculation formula]

Flg temp (FT) = CT + Cf X (CT-AT)

If the flange temperature exceeds the value preset

to

Flg temp Hi Alert Val or Flg temp Lo Alert Val,

an alert is generated.

NOTE

The flange temperature is calculated by the

calculation formula assumed that the capsule

part of EJX multivariable transmitter is heated up

or kept warm by an electrical heater or steam. In

the case of an atmosphere temperature or less,

the difference of temperature of approximately

3 to 4 °C, may occur because the amplifier

temperature becomes higher than the capsule

temperature.

4.2.3.1 Flg Temp Coef Setting

The value calculated according to the following

procedure is set to

Flg temp Coef.

• To enhance the calculation accuracy of the

flange temperature, measure the actual flange

temperature by using the temperature sensor

etc.

• Calculate the ratio of the capsule temperature

to the capsule temperature minus the amplifier

temperature from the two temperature values

measured by EJX multivariable transmitter.

• Derive the

Flg temp Coef from the measured

flange temperature and the ratio of the capsule

temperature to the amplifier temperature in

accordance with the following calculation

formula.

Flg temp Coef (Cf) =

(Actual measured value of Flange temperature) - CT

CT - AT

For DTM (HART 5), Flg Temp Coef parameter can

be set by using Tuning function.

By setting the measured value of flange

temperature, Flg Temp Coef is calculated by using

capsule temperature and amplifier temperature.

• Procedure to call up the

Tuning display.

Diag and Service → Advanced Diag Configuration →

Heat Trace → Tuning