Electromagnetic immunity declaration (eid) – ZOLL AED Pro Rev J User Manual
Page 95
Guidance and Manufacturer’s Declaration — Electromagnetic Emissions
9650-0350-01 Rev. J
ZOLL AED Pro Operator’s Guide
A–7
Electromagnetic Immunity Declaration (EID)
The ZOLL AED Pro device is intended for use in the electromagnetic environment specified
below. The customer or operator should ensure that the device is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment –
Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
Not applicable
±1 kV I/O
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV for common
mode
Not applicable
Not applicable
Voltage dips,
short
interruptions,
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% Ut (>95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 seconds
Note: Ut is the ac
mains voltage prior to
application of the test
level.
Not applicable
Not applicable
Not applicable
Not applicable
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.