Descriptions of AED Pro Configurable Settings
9650-0350-01 Rev. J
ZOLL AED Pro Operator’s Guide
C–3
Shock Energy Level and Charging
Adult First Shock Energy
Sets the energy level in joules for the first shock for an
adult patient.
•
120 J — default
•
150 J
•
200 J
Adult Second Shock Energy
Note: This value cannot be less than the value
selected for the first adult shock.
•
120 J
•
150 J — default
•
200 J
Adult Third Shock Energy
Note: This value cannot be less than the value
selected for the second adult shock.
•
120 J
•
150 J
•
200 J — default
Pediatric First Shock Energy
Sets the energy level in joules for the first shock for a
pediatric patient.
•
50 J — default
•
70 J
•
85 J
Pediatric Second Shock Energy
Note: This value cannot be less than the value
selected for the first pediatric shock.
•
50 J
•
70 J — default
•
85 J
Pediatric Third Shock Energy
Note: This value cannot be less than the value
selected for the second pediatric shock.
•
50 J
•
70 J
•
85 J — default
Enable Lay Rescuer
When this option is checked (On), the unit issues the
following voice and text prompts after completion of the
power-on self-test and entry into clinical mode:
STAY CALM
CALL FOR HELP
•
Off — default
•
On
Jump to Analysis
When this option is checked (On), the unit immediately
begins ECG analysis when pads are attached to the
patient, except during CPR periods.
Note: The Start with CPR option, if also checked (On),
overrides this setting.
•
On — default
•
Off
Unit OK
When this option is checked (On), the unit issues the
message Unit OK after a successful power-on self-test.
•
On
•
Off — default
Clinical Protocol Voice Prompts
Breathing
Allows you to select a check breathing prompt, based
on the guidelines you want to follow.
When disabled (Off), these prompts are not issued.
•
OFF — default
•
Open Airway, Check Breathing,
Give Two Breaths
•
Open Airway, Check Breathing
•
Give Two Breaths
Configurable Option
Possible Values