Applying Defibrillation Electrode Pads — Infant/Child pedi•padz II

WARNING!

ZOLL pedi•padz II electrodes are for pediatric patients only; these pads provide
defibrillation energy levels that might be inadequate for adult patients.

Procedure

To apply pedi•padz II defibrillation electrode pads:

Step

Action

1

Tear open the electrode package and unfold the inner
package to expose the electrodes.

2

Remove the round electrode from its backing material.

3

Position the electrode on the patient’s chest as shown.
Place your hand on the electrode edge. Using your other
hand, gently roll the electrode onto the patient’s chest,
pushing out air from beneath the electrode as you go.

4

Roll the patient onto his/her chest.

5

Remove the square electrode from its backing material.

6

Position the electrode on the patient’s back as shown.
Place your hand on the electrode’s edge. Using your
other hand, roll the electrode onto the patient’s skin,
pushing out air from beneath the electrode as you go.

7

Roll the patient onto his/her back, and follow the
AED Pro prompts.