beautypg.com

Ecg electrode placement – ZOLL AED Pro Rev J User Manual

Page 67

background image

ECG Electrode Placement

9650-0350-01 Rev. J

ZOLL AED Pro Operator’s Guide

5–3

ECG Electrode Placement

Before applying ECG electrodes, be sure to

Remove all clothing covering the patient’s chest.

Clip or shave any excessive hair to ensure proper adhesion of the electrodes.

Use alcohol to wash away any oil or dirt at the electrode site.

Dry any moisture at the electrode site.

Some skin abrasion is necessary to remove the top layer of dead skin cells and expose moist
living cells for better electrical contact. Proper skin preparation reduces baseline wander and
noise and provides a noise-free signal more quickly after electrode application.

Place ECG electrodes on the patient’s chest as shown in Table 5-1.

Table 5-1.

ECG Electrode Labeling and Placement

IEC

a

Label

a. International Electrotechnical Commission

AHA

b

Label

b. American Heart Association

Placement

R (red)

RA (white)

Patient’s right
midclavicular line, directly
below clavicle.

L (yellow)

LA (black)

Patient’s left midclavicular
line, directly below
clavicle.

F (green)

LL (red)

Between sixth and
seventh intercostal space
on the patient’s left
midclavicular line.

RA

LA

LL

R

L

F

3-lead

configuration

(AHA)

3-lead

configuration

(IEC)

ZM048004A

ZM048003A