beautypg.com

Ecg monitoring troubleshooting – ZOLL AED Pro Rev J User Manual

Page 84

background image

C

HAPTER

7

T

ROUBLESHOOTING

AND

M

AINTENANCE

7–4

www.zoll.com

9650-0350-01 Rev. J

ECG Monitoring Troubleshooting

Table 7-2 lists common issues with ECG monitoring and their associated corrective action.

Table 7-2.

ECG Monitoring Issues

Symptom

Corrective Action

Messages:
CHECK ECG ELECTRODES
ATTACH ECG ELECTRODES

Ensure that the ECG cable is connected to each
electrode and to the unit.

Ensure that the ECG electrodes are making good
contact with the patient and are not dried out.

Replace the ECG electrodes.

Replace the ECG cable.

Noisy ECG, artifact, or wandering
baseline.

Turn off nearby two-way radios and cell phones.

Before attaching electrodes, properly prepare the
patient’s skin (refer to “Applying ECG Electrodes” on
page 5-4).

Check for proper adhesion of the electrodes to the
patient.

Arrange the cable and leads so that they do not pull
on the electrodes or swing excessively.

Poor ECG signal level.

Replace the ECG electrodes, and change their
position on the patient.

Irregular heart rate.

Observe the patient’s ECG. Verify that the irregular
heart rate is not caused by noise, low amplitude
R waves, extra-systoles, or arrhythmias.

Replace the ECG electrodes, and change their
position on the patient.

Message:
PLUG IN DEFIB CABLE

ECG analysis detected a shockable rhythm.

Replace the ECG electrodes and cable with
defibrillation pads to deliver therapy.

ECG data recording stops.

If the unit is turned off and then turned on again within
10 seconds, ECG recording is interrupted.