Descriptions of aed pro configurable settings – ZOLL AED Pro Rev J User Manual
Page 110
C
ONFIGURABLE
S
ETTINGS
C–2 www.zoll.com
9650-0350-01 Rev. J
Descriptions of AED Pro Configurable Settings
Configurable Option
Possible Values
General Settings
Self-test Interval
Sets the period of time between automated self-tests in
standby state.
•
1 day — default
•
2 days
•
3 days
•
4 days
•
5 days
•
6 days
•
7 days
Line Frequency
Selects the AC power frequency to be filtered during
ECG monitoring.
•
60 Hz — default
•
50 Hz
Power Down Delay
Sets the time interval after which the unit powers off if
no patient connection is detected.
•
5 minutes
•
10 minutes — default
•
15 minutes
•
20 minutes
•
30 minutes
Audio Volume
Sets the volume level for voice messages and tones.
•
High — default
•
Medium
•
Low
Number of Patient Records
Sets the number of patients for whom information will
be stored in nonvolatile flash memory.
•
1
•
2
•
3
•
4 — default
Audio Recording Enabled
Enables the use of the Audio Recording Option
(requires the installation of an internal microphone
when the AED Pro unit is manufactured).
Note: You must set the Number of Patient Records to
1 before you can enable the Audio Recording
Option.
•
Disabled — default
•
Enabled
Monthly Test
When checked (On), the unit performs a monthly
full-energy self-test (120 joules).
•
On — default
•
Off
Daylight Savings
Allows the AED Pro’s 24-hour clock to automatically
adjust for Daylight Savings Time.
•
Manual — default
•
US Pre-2007
•
US
•
EU
•
EU Southern Hemisphere
Number of Shocks
Sets the number of shocks in the shock sequence to 1,
2, or 3 shocks.
•
1 Shock — default
•
2 Shocks
•
3 Shocks