C–2 www.zoll.com

9650-0350-01 Rev. J

Descriptions of AED Pro Configurable Settings

Configurable Option

Possible Values

General Settings

Self-test Interval

Sets the period of time between automated self-tests in
standby state.

1 day — default

2 days

3 days

4 days

5 days

6 days

7 days

Line Frequency

Selects the AC power frequency to be filtered during
ECG monitoring.

60 Hz — default

50 Hz

Power Down Delay

Sets the time interval after which the unit powers off if
no patient connection is detected.

5 minutes

10 minutes — default

15 minutes

20 minutes

30 minutes

Audio Volume

Sets the volume level for voice messages and tones.

High — default

Medium

Low

Number of Patient Records

Sets the number of patients for whom information will
be stored in nonvolatile flash memory.

1

2

3

4 — default

Audio Recording Enabled

Enables the use of the Audio Recording Option
(requires the installation of an internal microphone
when the AED Pro unit is manufactured).

Note: You must set the Number of Patient Records to

1 before you can enable the Audio Recording
Option.

Disabled — default

Enabled

Monthly Test

When checked (On), the unit performs a monthly
full-energy self-test (120 joules).

On — default

Off

Daylight Savings

Allows the AED Pro’s 24-hour clock to automatically
adjust for Daylight Savings Time.

Manual — default

US Pre-2007

US

EU

EU Southern Hemisphere

Number of Shocks

Sets the number of shocks in the shock sequence to 1,
2, or 3 shocks.

1 Shock — default

2 Shocks

3 Shocks