ZOLL AED Pro Rev J User Manual
Page 113

Descriptions of AED Pro Configurable Settings
9650-0350-01 Rev. J
ZOLL AED Pro Operator’s Guide
C–5
Post-Shock CPR Period
Sets the duration of the CPR period following the
delivery of one or more shocks.
When you select Extended, the device remains in the
CPR state indefinitely and displays the ANALYZE
softkey. You must press the ANALYZE softkey to move
to analysis.
•
30 seconds
•
60 seconds
•
90 seconds
•
120 seconds — default
•
150 seconds
•
180 seconds
•
Extended
Start with CPR (Semiautomatic Mode only)
When this option is checked (On), before beginning
ECG analysis, the unit prompts the rescuer to check the
patient’s pulse and perform a period of CPR if no pulse
is detected.
•
On
•
Off — default
Start with CPR Period
Sets the duration of the CPR period associated with the
Start with CPR option.
When you select Extended, the device remains in the
CPR state indefinately and displays the ANALYZE
softkey. You must press the ANALYZE softkey to move
to analysis
•
30 seconds
•
60 seconds
•
90 seconds
•
120 seconds — default
•
150 seconds
•
180 seconds
•
Extended
Analyze Key
Enables or disables the ANALYZE softkey during the
Post Shock and Post No Shock CPR periods. If the
ANALYZE softkey is enabled, it appears during both
the Post Shock and Post No Shock CPR periods.
•
On
•
Off — default
Fully Release
When checked (On), this text prompt reminds the
rescuer to lift the hands off the chest during CPR to
allow full recoil.
•
On
•
Off — default
CPR Prompts
Start with CPR Period
Start CPR
At the beginning of the first CPR period, if Start
with CPR Period is enabled, the AED Pro issues
one of the voice and text prompts shown in the
Possible Values column.
•
Open Airway, Check Breathing,
Check Pulse, If no Pulse Start
CPR
•
Open Airway, Check Breathing,
Check Circulation, If No
Circulation Start CPR
•
Start CPR — default
•
Start Compressions
•
If no Pulse, Start Compressions
•
If no Circulation, Start
Compressions
Configurable Option
Possible Values