Device Specifications

9650-0350-01 Rev. J

ZOLL AED Pro Operator’s Guide

A–3

Defibrillation electrode
pads

ZOLL single-use, pre-gelled electrode pads:

CPR-D•padz (includes CPR sensor)

CPR stat•padz (includes CPR sensor)

Adult stat•padz II

Pediatric pedi•padz II

Built-in defibrillator
self-test

Verifies proper charging and discharging of the defibrillator.

Defibrillation advisory

Evaluates electrode attachment and patient ECG to determine if
defibrillation is needed.

Shockable arrhythmia:

Ventricular fibrillation (VF) with amplitude greater than 100

V

Wide-complex ventricular tachycardia (VT)
Adult: greater than 150 beats per minute
Pediatric: greater than 200 beats per minute

Valid patient impedance
range

10

 to 300

CPR Monitoring

Compression depth

0.75 to 3 inches ±0.25 inches
1.9 to 7.6 cm ±0.6 cm

Compression rate

50 to 150 compressions per minute

ECG Monitoring

Input protection

Fully defibrillation-protected

Implanted pacemaker
pulses detected

The AED Pro does not reject implanted pacemaker pulses.

Bandwidth

1.4 to 22 Hz with defibrillation electrode cable

1.4 to 22 Hz (default) with AED Pro ECG cable;
0.7 to 30 Hz as a configurable option

ECG lead

Lead II

ECG amplitude range

±5 mV

Heart rate range

30 to 300 beats per minute

Heart rate accuracy

±5 beats per minute

Heart rate resolution

1 beat per minute

Heart rate alert

Configurable low heart rate limit in the range 30 to 100 beats per
minute

Off

Heart rate alarm

Voice and text prompt indicates activated/deactivated status.
Tachycardia: 250 bpm
Bradycardia: 30 bpm - 100 bpm, user selectable

Tall T-wave rejection

9 mm (ECG scale at x1)
0.9 mV