ZOLL AED Pro Rev J User Manual
Page 112

C
ONFIGURABLE
S
ETTINGS
C–4 www.zoll.com
9650-0350-01 Rev. J
Breathing Prompt Delay
Sets the period over which the unit issues the check
breathing prompts.
•
2 seconds
•
4 seconds
•
6 seconds — default
•
9 seconds
•
15 seconds
•
20 seconds
Responsiveness/Patient
Selects the wording for this prompt.
•
OFF — default
•
Check Responsiveness
•
Check Patient
Responsiveness/Patient Prompt Delay
Sets the duration before the next prompt or disables
this prompt.
•
2 seconds
•
4 seconds — default
•
6 seconds
•
8 seconds
Circulation/Pulse
Selects the wording for this and related prompts.
•
OFF — default
•
Check Circulation
•
Check Pulse
Circulation/Pulse Prompt Delay
Sets the duration before the next prompt or disables
this prompt.
•
2 seconds
•
5 seconds
•
7 seconds
•
10 seconds — default
•
15 seconds
•
20 seconds
Press Treatment/Shock
Selects the wording for this and related prompts.
•
Press Flashing Treatment Button
•
Press Flashing Shock Button —
default
CPR
CPR Monitoring
When this option is checked (On) and CPR-D•padz or
CPR stat•padz are attached, the unit performs CPR
monitoring and prompts the rescuer.
•
On — default
•
Off
No-Shock CPR Period
Sets the duration of the CPR period following a
no-shock result for the first analysis in a stack.
When you select Extended, the device remains in the
CPR state indefinitely and displays the ANALYZE
softkey. You must press the ANALYZE softkey to move
to analysis.
•
30 seconds
•
60 seconds
•
90 seconds
•
120 seconds — default
•
150 seconds
•
180 seconds
•
Extended
Cycle through CPR
When this option is checked (On) and pads are
detached from the patient for more than one and
one-half minutes, the unit prompts the rescuer to
perform CPR.
•
On — default
•
Off
Configurable Option
Possible Values