ZOLL AED Pro Rev J User Manual
Page 116
C
ONFIGURABLE
S
ETTINGS
C–8 www.zoll.com
9650-0350-01 Rev. J
ECG
Display ECG Waveform
When this option is checked (On), the unit displays the
patient’s ECG rhythm.
•
On — default
•
Off
Monitor BW Filter Cutoff
Selects the bandwidth filter to be used during ECG
monitoring.
•
1.35 – 22 Hz — default
•
0.7 – 30 Hz
ECG Analysis During CPR
When this option is checked (On), background ECG
analysis begins after the start of a CPR period. If a
shockable rhythm is detected during a 12-second
period in which CPR is not performed, the unit prompts
the rescuer to stop CPR and stand clear. It then initiates
ECG analysis.
When this option is not checked (Off), the unit does not
perform ECG analysis during CPR periods.
•
Off — default
•
On
Use Analysis After
When the ECG Analysis During CPR option is checked
(On), this option specifies the number of seconds to
delay background ECG analysis after a no-shock or
post-shock CPR period begins.
•
0 seconds — default
•
15 seconds
•
30 seconds
•
45 seconds
•
60 seconds
•
120 seconds
•
180 seconds
See-Thru CPR
When this option is checked (On), See-Thru CPR is
enabled for the AED Pro (see page 4-5 for information
on using See-Thru CPR.).
•
On
•
Off — default
Archive ECG AF Data
When this option is checked (On), the See-Thru CPR
adaptive filter information will be saved.
•
On
•
Off — default
Monitoring
ECG Monitoring Modes
When this option is checked (On), the unit switches to
ECG monitoring mode when an ECG cable is attached
or when the left softkey is pressed and held for at least
5 seconds in semiautomatic mode.
When this option is not checked (Off), the unit does not
support ECG monitoring mode. If an ECG cable is
attached, the unit prompts the rescuer to plug in a
defibrillation cable.
•
On — default
•
Off
Display Heart Rate
When this option is checked (On), the unit displays the
patient heart rate in manual mode.
•
On — default
•
Off
Configurable Option
Possible Values