C

ONFIGURABLE

S

ETTINGS

C–8 www.zoll.com

9650-0350-01 Rev. J

ECG

Display ECG Waveform

When this option is checked (On), the unit displays the
patient’s ECG rhythm.

On — default

Off

Monitor BW Filter Cutoff

Selects the bandwidth filter to be used during ECG
monitoring.

1.35 – 22 Hz — default

0.7 – 30 Hz

ECG Analysis During CPR

When this option is checked (On), background ECG
analysis begins after the start of a CPR period. If a
shockable rhythm is detected during a 12-second
period in which CPR is not performed, the unit prompts
the rescuer to stop CPR and stand clear. It then initiates
ECG analysis.

When this option is not checked (Off), the unit does not
perform ECG analysis during CPR periods.

Off — default

On

Use Analysis After

When the ECG Analysis During CPR option is checked
(On), this option specifies the number of seconds to
delay background ECG analysis after a no-shock or
post-shock CPR period begins.

0 seconds — default

15 seconds

30 seconds

45 seconds

60 seconds

120 seconds

180 seconds

See-Thru CPR

When this option is checked (On), See-Thru CPR is
enabled for the AED Pro (see page 4-5 for information
on using See-Thru CPR.).

On

Off — default

Archive ECG AF Data

When this option is checked (On), the See-Thru CPR
adaptive filter information will be saved.

On

Off — default

Monitoring

ECG Monitoring Modes

When this option is checked (On), the unit switches to
ECG monitoring mode when an ECG cable is attached
or when the left softkey is pressed and held for at least
5 seconds in semiautomatic mode.

When this option is not checked (Off), the unit does not
support ECG monitoring mode. If an ECG cable is
attached, the unit prompts the rescuer to plug in a
defibrillation cable.

On — default

Off

Display Heart Rate

When this option is checked (On), the unit displays the
patient heart rate in manual mode.

On — default

Off

Configurable Option

Possible Values