Applying Defibrillation Electrode Pads

9650-0350-01 Rev. J

ZOLL AED Pro Operator’s Guide

3–7

Applying Defibrillation Electrode Pads — Adult stat•padz II

WARNING!

ZOLL stat•padz II electrodes are for adult patients only; do not use them on patients
under 8 years of age.

To apply stat•padz II defibrillation electrode pads:

Step

Action

1

Tear open the electrode package and unfold the inner
package to expose the electrodes.

2

Remove the square electrode from its backing material,
and place it on the patient’s upper right chest as shown.

3

Place your hand on the electrode edge. Using your
other hand, gently roll the electrode onto the patient’s
chest, pushing out air from beneath the electrode as
you go.

4

Remove the round electrode from its backing material,
and place it on the patient’s lower left chest as shown.

For female
patients, place the
electrode under
the patient’s left
breast.

5

Place your hand on the electrode’s edge. Using your
other hand, roll the electrode onto the patient’s skin,
pushing out air from beneath the electrode as you go.

6

Follow the AED Pro prompts.