ZOLL AED Pro Rev J User Manual
Page 114
C
ONFIGURABLE
S
ETTINGS
C–6 www.zoll.com
9650-0350-01 Rev. J
Continue CPR
When you select one of the prompts in the
Possible Values column, that prompt will be
repeated once every 10 seconds (default) during
the CPR period if compression stops while in the
CPR state
When this option is set to Off, no prompts will be
issued.
•
OFF — default
•
If no Circulation, Continue CPR
•
If no Pulse, Continue CPR
•
Continue CPR
•
Continue Compressions
•
If no Pulse, Continue
Compressions
•
If no Circulation, Continue
Compressions
Post Shock CPR Period
Start CPR
At the beginning of the CPR period, after the
ECG analysis, the AED Pro issues one of the
voice and text prompts shown in the Possible
Values column.
•
Open Airway, Check Breathing,
Check Pulse, If no Pulse Start
CPR
•
Open Airway, Check Breathing,
Check Circulation, If No
Circulation Start CPR
•
Start CPR — default
•
Start Compressions
•
If no Pulse, Start Compressions
•
If no Circulation, Start
Compressions
Continue CPR
When you select one of the prompts in the
Possible Values column, that prompt will be
repeated once every 10 seconds (default) during
the CPR period if compression stops while in the
CPR state.
When this option is set to Off, no prompts will be
issued.
•
OFF — default
•
If no Circulation, Continue CPR
•
If no Pulse, Continue CPR
•
Continue CPR
•
Continue Compressions
•
If no Pulse, Continue
Compressions
•
If no Circulation, Continue
Compressions
Post No Shock CPR Period
Start CPR
At the beginning of the CPR period after the
ECG analysis, the AED Pro issues one of the
voice and text prompts shown in the Possible
Values column.
•
Open Airway, Check Breathing,
Check Pulse, If no Pulse Start
CPR
•
Open Airway, Check Breathing,
Check Circulation, If No
Circulation Start CPR
•
Start CPR — default
•
Start Compressions
•
If no Pulse, Start Compressions
•
If no Circulation, Start
Compressions
Configurable Option
Possible Values