The Front Panel
9650-0350-01 Rev. J
ZOLL AED Pro Operator’s Guide
1–9
Table 1-1.
AED Pro Front Panel Features
Feature
Description
Shock button
When the defibrillator is fully charged and ready, the Shock button
repeatedly flashes. To deliver a shock, press and hold the button.
On/Off button
To start the unit, press this button and release it within 5 seconds.
To start the unit in nonrescue mode, press and hold this button for
more than 5 seconds.
To turn the unit off and place it in standby state, press and hold this
button for 1 second.
Ready indicator
Shows the status of the unit, based on
its last self-test.
A green check indicates the unit is
ready for use.
A red “X” indicates the unit is not ready
for use.
IrDA port
Provides a way to connect the unit to an external device for
transferring patient data, unit status information, or configuration
information.
Speaker
Issues voice prompts and alerts.
Battery compartment
Holds the battery pack.
Battery compartment latch
Provides access to the battery compartment.
Softkeys
Directly below the display, two unlabeled buttons control various
functions depending on the operating mode. Labels for the softkeys
appear at the bottom of the display above each softkey to indicate
its function.
Patient cable connector
Used for plugging in defibrillation electrodes or an AED Pro ECG
cable.
USB connector
Allows you to download patient data to a USB device.
Microphone (optional)
Allows the AED Pro Unit to record audio rescue data. Only AED Pro
units that have been ordered with the Audio Recording Option have
a microphone installed.
ZM040010A
This connector is a defibrillator-proof type BF patient
connection.