Digital descriptor box, Timebase descriptor box, Trigger descriptor box – Teledyne LeCroy HDO6000 Operators Manual User Manual
Page 20
HDO6000 High Definition Oscilloscope
Preprocessing Symbols on Descriptor Boxes
Pre-Processing Type
Long Form
Short Form
Sin X Interpolation
SINX
S
Averaging
AVG
A
Inversion
INV
I
Deskew
DSQ
DQ
Coupling
DC50, DC1M or AC1M
D50, D1M, or A1
Ground
GND
G
Bandwidth Limiting
BWL
B
Similar descriptor boxes appear for zoom (Zx), math (Fx), and memory (Mx) traces. These descriptor
boxes show any Horizontal scaling that differs from the signal Timebase.
Digital Descriptor Box
Digital descriptor boxes appear whenever a digital line group is enabled on a mixed-
signal model oscilloscope. Like Channel descriptors, they are numbered 1-4
corresponding to one of the four line groups. They show the number of digital lines in
the group, the digital sample rate, and the digital memory.
Timebase Descriptor Box
The TimeBase descriptor box shows: (clockwise from top right) Trigger Delay
(position), Time/div, Sample Rate, Number of Samples, and Sampling Mode (blank
when in real-time mode).
Trigger Descriptor Box
Trigger descriptor box shows: (clockwise from top right) Trigger Source and Coupling,
Trigger Level (V), Slope, Trigger Type, Trigger Mode.
Setup information for Horizontal cursors, including the time between cursors and the
frequency, is shown beneath the TimeBase and
Trigger descriptor boxes. See the
section for more information.
12
922499 Rev B