beautypg.com

Certifications, Emc compliance – Teledyne LeCroy HDO6000 Operators Manual User Manual

Page 138

background image

HDO6000 High Definition Oscilloscope

Certifications

This section certifies the instrument’s Electromagnetic Compatibility (EMC), Safety and Environmental
compliances.

EMC Compliance

EC Declaration of Conformity- EMC

The oscilloscope meets intent of EC Directive 2004/108/EC for Electromagnetic Compatibility.
Compliance was demonstrated to the following specifications as listed in the Official Journal of the
European Communities:

EN 61326-1:2006, EN 61326-2-1:2006 EMC requirements for electrical equipment for measurement,
control, and laboratory use. 1

E

LECTROMAGNETIC

E

MISSIONS

:

CISPR 11:2003, Radiated and Conducted Emissions Group 1, Class A 2 3

EN 61000-3-2:2006 Harmonic Current Emissions, Class A

EN 61000-3-3/A2:2005 Voltage Fluctuations and Flickers, Pst = 1

E

LECTROMAGNETIC

I

MMUNITY

:

EN 61000-4-2:2001 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV vertical/horizontal coupling
planes 4

EN 61000-4-3:2006 RF Radiated Electromagnetic Field, 3 V/m, 80-1000 MHz; 3 V/m, 1400 MHz - 2 GHz; 1
V/m, 2 GHz - 2.7 GHz 4

EN 61000-4-4:2004 Electrical Fast Transient/Burst, 1 kV on power supply lines, 0.5 kV on I/O signal data
and control lines 4

EN 61000-4-5:2006 Power line Surge, 1 kV AC Mains, L-N, L-PE, N-PE 4

EN 61000-4-6:2007 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz - 80 MHz 4

EN 61000-4-11:2004 Mains Dips and Interruptions, 0%/1 cycle, 70%/25 cycles, 0%/250 cycles 4 5

1 To ensure compliance with all applicable EMC standards, high quality shielded interface cables should be used.
2 Emissions which exceed the levels required by this standard may occur when the oscilloscope is connected to a test object.
3 This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
4 Meets Performance Criteria “B” limits of the respective standard: during the disturbance, product undergoes a temporary
degradation or loss of function or performance which is self-recoverable.
5 Performance Criteria “C” applied for 70%/25 cycle voltage dips and for 0%/250 cycle voltage interruption test levels per
EN61000-4-11.

E

UROPEAN

C

ONTACT

:

Teledyne LeCroy Europe GmbH
Waldhofer Str 104
D-69123 Heidelberg
Germany
Tel: (49) 6221 82700

130

922499 Rev B