Operator's Manual
Set Up Remote Control
Contact your Network Administrator to connect the oscilloscope to your LAN. Use a USB cable to connect
the oscilloscope directly to a PC.
1. From the menu bar, choose Utilities > Utilities Setup, then touch the Remote tab.
2. On the Remote dialog, make a Control From selection.
3. If you are using TCPIP or LXI, touch the Net Connections button. The Windows Network Connections
window opens for you to select a network.
4. If you are using TCPIP and wish to restrict control of the oscilloscope to specific network clients,
touch Yes. Enter the IP addresses or DNS names of the authorized controllers in a comma-delimited
list.
Configure the Remote Control Assistant Event Log
The Remote Control Assistant monitors communication between the controller and oscilloscope when
you are operating the instrument remotely. You can log all events or errors only. The log can be output to
an ASCII file and is invaluable when you are creating and debugging remote control programs.
1. From the menu bar, choose Utilities > Utilities Setup, then touch the Remote tab.
2. Under Remote Control Assistant, touch Log Mode and choose Off, Errors Only, or Full Dialog.
3. To always clear the log at startup, check Reset to Errors Only and clear at startup.
Export Contents of the Event Log
1. From the menu bar, choose Utilities > Utilities Setup, then touch the Remote tab.
2. Touch the Show Remote Control Log button. The Event Logs pop-up is shown.
3. Enter a log file name in DestFilename, or touch Browse and navigate to a file.
NOTE: New contents will overwrite the existing content; it is not appended.
4. Touch Export to Text File.
922499 Rev B
99