Maintenance and Calibration

Model 3500

Teledyne Analytical Instruments

72

absorption by the calibration tube and the connection tubes are
negligible.

6.2.1 Calibration Procedure

To calibrate the system, follow the procedure listed below:

1. Turn off the system power supply.
2. Loosen the mounting nut and take off the transmitter unit and the

receiver unit (Figure 6.2). Check the optical parts for cracks, dust
or other contaminants. If damage or contamination is evident
refer to section 6.1.2 to clean and maintain the optical parts.

Figure 6.2 Removing the Transmitter Receiver Units

3. Mount the transmitter and receiver units to the flanges at the

each end of the calibration devices, then and fasten the locknut
(Figure 6.3)