Laser Gas Analyzer

Installation

Teledyne

Analytical

Instruments

17

7. Repeat step 4 to tighten 4 fastening screws on the other

instrument flange.

3.3.4 Installing the Transmitter Unit and the Receiver Unit

Mount the transmitter unit on one of the instrument flanges as

shown in Figure 3.1. Tighten it with the mounting nut, then tighten the
fastening screw. Mount the receiver unit the same way.

3.4 Central Processing Unit Installation

The central processing unit is a wall-mountable enclosure that is

hung on the wall of the instrument control center or in an analysis room
in the field. There are four pothooks on the housing for this purpose.
Secure the unit case to the wall using 4 M6 screws. Refer to Figure 3.5
for the case dimensions. The central processing unit should be hung at a
proper height for comfortable operation.

Figure 3.5 Dimension of the Central Processing Unit (in mm).

To avoid turning on the power supply inadvertently and possibly

causing harm to operators in the field, the central processing unit is