Operation Model

3500

Teledyne Analytical Instruments

46

measurement result.

4.3.4 Cali.

Cali. main menu is third from left on MainMenu. Pressing “SET”,

the system prompts for the password, if the password is entered
correctly, the system enters Cali. main menu. (Figure 4.23)

Note: In error mode, the system is in Error Protection Menu (see

Figure 4.39), and the Cali. menu is not accessible.

Figure 4.23 Cali. Main Menu


From this menu, the operator can perform the following operations:

• Adjust zero point: Fill the calibration tube with standard gas

(such as high purity nitrogen) with 0% of the gas under test, and
zero calibrate.

• Calibrate coefficient: Fill the calibration tube with standard gas

(such as high purity Nitrogen) with known concentration of the
gas under test, and perform a span calibration operation.

Note: Perform the zero calibration before doing the span

calibration.