Laser Gas Analyzer

Alarm Messages

Teledyne

Analytical

Instruments

63

Table 5.2: LCD Display Alarm Messages

Alarm

Code on

LCD

Possible

Cause of the

alarm

Relay Status

a. Warning Alarm

b. Error Alarm

c. Conc. Alarm

(O=Open

C=Closed)

4-20mA

Output

Trouble shooting

02
03
11
13
15
17
18
57
58

Model 3500S
has internal
system error

a=O b=O c=C

2mA

Write down the alarm code;
power off the system; reboot
the system in an hour; If the
same alarm persists, contact
technical support center.

05/06

The temperature
inside the Model
3500S central
processing unit
is too high/low.

a=O b=O c=C

2mA

Power off the system; check
whether the environment
temperature inside/outside of
the central processing unit is
abnormal; reboot the system
after making sure that the
temperature is normal. If the
same alarm persists, contact
the technical support center.

08/09

The laser
temperature is
too high/low.

a=O b=O c=C

2mA

Power off the system; reboot
the system in an hour; If the
same alarm persists, contact
the technical support center.

20

Background
radiation is too
strong during
system self test.

a=O b=O c=C

2mA

Power off the system; reboot
the system in an hour; If the
same alarm persists, contact
the technical support center.