Laser Gas Analyzer

Operation

Teledyne

Analytical

Instruments

49

The functions of the buttons on Cali.Co sub menu are:

• Cali.Len: Sets the optical path length of the calibration tube,

manual.

• Cali.Temp: Sets the temperature of the standard gas used for

calibration, manual or from a measurement. When “Measure” is
chosen, the system enters Measured Temperature Set menu.

• Cali.Pres: Sets the pressure of the standard gas used for

calibration, manual or from a measurement. When “Measure” is
chosen, the system enters Measured Pressure Set menu.

• Cali.Con: Enters the concentration value of the standard gas used

for calibration. This value is used to calculate calibration
coefficients.

• Pre.Cali: After inputting all the parameters above, press Pre.Cali

button to accept them, and the system enters the Calibration
Confirmation menu (Figure 4.27).

Note: The system accepts the above input parameters only after

Pre.Cali is pressed.

Figure 4.27 Calibration confirmation menu