Laser Gas Analyzer

Operation

Teledyne

Analytical

Instruments

47

• Backup coefficient: Save current calibration coefficients as

backup. The operator can choose to backup any set of calibration
coefficients they are satisfied with. The Model 3500 has
calibration coefficients backed up before shipping.

• Restore coefficient: When the users are not happy with the

current calibration result, and are unable to get good calibration
results, they can restore the old more satisfying backup
calibration coefficients.

4.3.4.1

A

DJUST

Z

ERO

:

Zero: Press “SET” when the cursor is on “Zero” button, the system
enters the Pre zeroing menu (Figure 4.24).

Figure 4.24 Pre zeroing sub menu


Choose “Accept” to accept the parameters, and the system starts

zeroing (Figure 4.25). The zeroing gauge bar shows the progress of the
process.