beautypg.com

Teledyne LGA-3500 - Laser Gas Analysis system User Manual

Page 35

background image

Laser Gas Analyzer

Installation

Teledyne

Analytical

Instruments

21

Figure 3.8: Electric Connection Ports on the Central Processing

Unit for Analysis Systems Using Fiber Optic Pigtail

Table 3.1 Electric connections & definitions of Input, Output Signals of
Central Processing Unit

Socket No. On

Motherboard

Name

31

5V Power Supply

32

DGND Power Supply Ground

33 RS232-RXD

34 RS232-GND

35 RS232-TXD

36

4-20mA Concentration Output (-)

37

4-20mA Concentration Output (+)

38

4-20mA Pressure Input (-)

39

4-20mA Pressure Input (+)

40

4-20mA Pressure Input24V Power Supply

41

4-20mA Temperature Input (-)

42

4-20mA Temperature Input (+)

43

4-20mA Temperature Input 24V Power Supply

44

Warning Alarm Relay 1

1

. Grounding pole

2

. RS232 connection port

3

. Cable port between central

processing unit and the
probe

4

. 4-20mA connection port

5

. Relay connection port

6

. Fuse socket

7 Power supply port