Laser Gas Analyzer

Operation

Teledyne

Analytical

Instruments

35

Users can set the unit for concentration, length, temperature and

pressure by pressing “SET” in different cursor locations. For example,
press “SET” when the cursor is on Con. Unit, the system enters the
following menu:

Figure 4.11 Set Con. Unit menu


The following units for concentration, length, temperature and

pressure are available:

• Con. Unit: ppm, %, g/m

3

, g/Nm

3

, mg/m

3

, mg/Nm

3

• Len. Unit: m, cm, ft, inch
• Temp. Unit: °F, K, °C
• Pres. Unit: bar, Pa, mbar, psi, Mpa


Six units are provided in the Con. Unit menu to choose from: ppm,

%, g/m

3

, g/Nm

3

, mg/m

3

, and mg/Nm

3

. Units in ppm and % are

commonly used for volume percentage concentration while g/m

3

and

mg/m

3

are mass volume ratio concentration units related to volume

percentage concentration through the ideal gas equation

.

The units

g/Nm

3

and mg/Nm

3

are mass volume ratio concentration units (standard

state , pressure: 1.01325Bar, temperature: 273.1K).