beautypg.com

3 lcd alarm message display – Teledyne LGA-3500 - Laser Gas Analysis system User Manual

Page 76

background image

Alarm Messages

Model 3500

Teledyne Analytical Instruments

62

Table 5.1 Relays and output status for Various Operation Modes

Relay

System Working

Status

LCD

Status

Bar

LCD

Alarm

Icon

(Graphics

Mode)

Warning

Alarm

(Relay_1)

Error Alarm

(Relay_2)

Concentration

Alarm

(Relay_3)

4-

20mA

Output

Power-off

No

No

Open = Alarm

Open = Alarm

Open (Alarm)

0mA

Start-up Mode

No No

Closed = No

Alarm

Closed = No

Alarm

Closed = No

Alarm

0mA

Normal

Status

Normal No

Closed = No

Alarm

Closed = No

Alarm

Depending on

gas

concentration

4-

20mA

Normal

Mode

Warning

Status

Alarm
Message
Code

Open = Alarm

Closed = No

Alarm

Depending on

gas

concentration

4-

20mA

Error Mode

Alarm

Message

Code

Open = Alarm

Open = Alarm

Closed = No

Alarm

2mA

5.3 LCD Alarm Message Display

When the Model 3500 enters an alarm condition an alarm message

is displayed on the LCD screen. Detailed descriptions and solutions of
these messages are listed in Table5.2.

.