Operation Model

3500

Teledyne Analytical Instruments

30

Figure 4.7 Numerical Value Mode of the MainMenu


The meaning of each item on the menu is listed below:

• Current concentration: the gas concentration value just measured
• Average concentration: gas concentration value after digital

filtering. It can alleviate the influence caused by the uncertainty
of gas flow.

• Standard dev: standard deviation of the measured gas

concentration.

• Gas temperature: the temperature of the gas under test.
• Gas pressure: the pressure of the gas under test.
• Ave. times: Number of times used in digital filtering.
• Trans: Transmittance of laser beam in the environment under

test. If the optical windows in the transmitter and the receiver are
contaminated by dust in the environment under test, or if the
optical path between the transmitter and the receiver deviates
from the optimal position, the transmittance decreases.

• Con. output--4mA: gas concentration value corresponding to

4mA of the 4-20mA gas concentration output.