Laser Gas Analyzer

Operation

Teledyne

Analytical

Instruments

31

• Con. output--20mA: gas concentration value corresponding to

20mA of the 4-20mA gas concentration output.

• System time: the time of system clock.
• System date: the date of system clock.

Note: The Model 3500 outputs average gas concentration at the

4-20mA gas concentration output port.

The computation method for average concentration and standard

deviation is described below:

Average Concentration

old

cur

new

g

N

g

N

g

⎛ −

+

=

1

1

1

Where:

new

g

——

Average concentration value

cur

g

——

Current concentration value

old

g

—— the average concentration value before adding the

latest measurement

cur

g

stored in the system memory

N

——Ave. times

The method is in equivalent to having the current concentration

value pass through a low-pass filter with a cut-off frequency at

τ

π

N

f

2

1

=

,

where

τ

is the time the system needs to finish one

measurement. Typically, the Model 3500 series instruments have

τ

set

at

∼1 second at the factory. However, it can also be set at any value in

the range of 0.1~1 second upon customer request to suit particular
applications that require faster response time. Contact technical support
for more information.

Normally, every time the system finishes one measurement cycle,

the LCD display and output at the 4-20mA output port refreshes.

Standard Deviation