Laser Gas Analyzer

Operation

Teledyne

Analytical

Instruments

43

Figure 4.19: Measured Pressure Menu


Items displayed on the Measured Pressure screen (Figure 4.19) are:

• Pres. input--4mA: The gas pressure input is taken from a

pressure measurement. It is input to correspond with a 4-20mA
output signal. Pres. input--4mA is the gas pressure corresponding
to 4mA.

• Pres. input--20mA: This is also taken from a measurement. It is

the gas pressure corresponding to 20mA.

The buttons on this menu refer to:

4mA:

Set the corresponding gas pressure when the pressure
input signal is 4mA.

20mA:

Set the corresponding gas pressure when the pressure
input signal is 20mA.

3. Concentration Output Menu

Items displayed on the Concentration Output screen (Figure 4.20)

are:

• Con. output--4mA: The measured gas concentration output is in

the form of 4-20mA. Con. output--4mA is the gas concentration
value corresponding to 4mA.