Performing a calibration, Performing a 2-point calibration – PASCO SPARK Science Learning System User's Guide User Manual

Page 22

background image

14

S

ETTING

UP

AN

EXPERIMENT

C

ALIBRATING

SENSORS

5.

Touch the Calibration Type: box and select a calibration type.

See the instructions provided with your sensor for information about what
type of calibration is appropriate for your sensor.

6.

Touch NEXT.

The Calibrate Sensor: Enter Values screen opens.

Now that you have selected a measurement, you can perform a calibration as
described in the next task.

Performing a calibration

After you have selected a measurement to be calibrated and a calibration type
(see previous task), you are ready to perform a calibration.

Depending on the calibration type that you have selected, perform a 2-point
calibration, a 1-point offset-only calibration, or a 1-point slope-only calibration.

Performing a 2-point calibration

In the Calibrate Sensor: Enter Values screen, complete these steps:

1.

Apply a known quantity to the sensor.

For example, place a pH probe in a pH 4 buffer solution.

2.

Under Calibration Point 1 touch the Standard Value: box and enter
the known value.

For example, enter the known pH of the buffer solution.

3.

Under Calibration Point 1 touch Read From Sensor.

The value measured by the sensor is transferred to the Sensor Value: box.

4.

Apply a different known quantity to the sensor.

For example, place the pH probe in a pH 7 buffer solution.

5.

Under Calibration Point 2 touch the Standard Value: box and enter
the known value.

For example, enter the known pH of the buffer solution.

6.

Under Calibration Point 2 touch Read From Sensor.

The value measured by the sensor is transferred to the Sensor Value: box.

7.

Touch OK.

Calibration is complete.