Multichannel Systems MEA Manual User Manual
Page 93
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11
[email protected]
www.multichannelsystems.com
© 2013 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.
pMEA32S12 Layout 3
Technical Specifications pMEA32S12 Layout 3
Temperature compatibility
10 - 50 °C
Dimension (W x D x H)
49 mm x 25 mm x 1.8 mm
Base material
Polyimide foil on ceramic carrier with perforation
Perforation:
Total area of holes
0.8 mm
Diameter of the holes
90, 75, 50, 30, 20 μm
Track
material
Ti
(Titanium)
Contact
pads
TiN
(Titanium
nitride)
Electrode
diameter
30
μm (recording electrodes) 50 μm (stimulation electrodes)
Interelectrode distance (centre to centre)
90 μm and 150 μm (recording electrodes),
90 μm and 100 μm (stimulation electrodes)
Electrode
height Planar
Electrode
type
TiN
(Titanium
nitride)
electrodes
Isolation
type
Polyimide
foil
Electrode impedance
Approximately 30 - 50 k
Electrode layout grid
1 x 10 + 1 x 12 + 1 x 10 (recording electrodes),
3
x
4
(stimulation
electrodes)
Number of recording electrodes
32
Number of stimulation electrodes
12
Number of reference electrodes
1 internal reference electrode
MC_Rack “Source layout”
Configuration (MEA2100-32)
in “Data Source Setup”
1 dimensional, no digital channel (USB-MEA32-STIM4)
Channel
map
pMEA-32S12-L3_12x3.cmp
Cleaning Rinse
with
distilled
water.
Do
not
use
ultrasonic
bath!
Do not autoclave or sterilize pMEAs by heat.
These MEA types are not heat-stable and will be irreversibly damaged!
pMEA perfusion chamber
(w/o) Without ring
(gr) Glass ring: ID +/- 19 mm, OD 24 mm, height 6 / 12 mm
2
Perforated MEA for use with MEA2100-32- and USB-MEA32-STIM4-System