FlexMEA72
Technical Specifications FlexMEA72
Temperature compatibility
10 - 40 °C
Dimension (W x D) of the FlexMEA72
42 mm x 12 mm
Thickness of the electrode field
12 μm
Weight
<
1
g
Base
material
Polyimide
2611
foil
Contact pads and track material
Gold
Electrode
diameter
100
μm
Interelectrode distance (centre to centre)
625 to 750 μm
Perforation
Diameters
of
the
holes
100
μm
Electrode
height Planar
Electrode
type
TiN
electrodes
(Titanium
nitride)
Isolation
type
Polyimide
2611
foil
Electrode
impedance
Approximately
50
k
Electrode
layout
grid
9
x
8
Number of recording electrodes
64
Number of reference electrodes
4 internal reference electrodes
Number
of
ground
electrodes
4
MC_Rack
Version
4.1.1
and
higher
Data Source Setup
Source Layout: Configuration, Total Number of Channels: 64,
Amplifier:
FA64I/S,
MEA:
FlexMEA72.
Channel map
Display: Tab Layout > Default Map (FlexMEA72.cmp).
Cleaning
Rinse with distilled water, optional with ethanol 70%.
Do not autoclave or sterilize FlexMEAs by heat. These MEA types are not heat-stable
and will be irreversibly damaged! Please do not use an ultrasonic bath for cleaning.
Flexible Microelectrode Array with 72 electrodes for use via ADPT-FM-72 adapter with
two 32-Channel Miniature Preamplifier MPA32I for in vivo and in vitro applications.