Ecoflexmea36 – Multichannel Systems MEA Manual User Manual
Page 134
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11
[email protected]
www.multichannelsystems.com
© 201
3 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.
EcoFlexMEA36
Technical Specifications EcoFlexMEA36
Temperature
compatibility
10
-
125
°C
Dimension (W x D)
37 mm x 30 mm
Thickness of the electrode field
50 μm
Weight
<
10
g
Base
material
Polyimide
(Kapton)
Contact pads and track material
Gold
Electrode
diameter
50
μm
Interelectrode distance (centre to centre)
300 μm
Electrode
height Planar
Electrode
type
Gold
electrodes
Isolation
type
Polyimide
(Kapton)
Electrode
impedance
Approximately
50
k
Electrode layout grid
6 x 6
Number of electrodes
36
Reference electrodes
2 internal reference electrodes
Ground
electrodes
2
ground
electrodes
MC_Rack
Source layout in “Data Source Setup”
1 dimensional, 32 channels, no digital channel
Channel
map
EcoFlexMEA36.cmp
Cleaning
Rinse with distilled water. EcoFlexMEAs made from
Polyimide
(Kapton)
are
heat
stable
and
autoclavable.
Electrode field:
1800 x 1800 μm,
32 recording electrodes,
2 reference electrodes,
2 ground electrodes
Connect the EcoFlexMEA36 directly to a 32-Channel Miniature
Preamplifier. Insert the EcoFlexMEA36 into the MPA32I with
the electrode field up, when the backside of the MPA32I with
the screws is upside down. The additional connector can be used
for connecting a silver pellet or a silver wire for grounding the bath.
Flexible Microelectrode Array with 36 electrodes for use with
32-Channel Miniature Preamplifier MPA32I for in vivo or in vitro applications.
37 mm
9 mm
28 mm
30 mm
23 mm
6 mm
14 mm
Upside
of MPA32I
Electrode
field