Standard 60mea, Standard electrode layout grid 8 x 8 – Multichannel Systems MEA Manual User Manual
Page 68
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11
[email protected]
www.multichannelsystems.com
© 2014 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.
Standard 60MEA
60MEA200/10iR-ITO, 60MEA200/30iR-ITO,
60MEA100/10iR-ITO, 60MEA100/10-ITO,
60MEA100/10iR-Ti, 60MEA200/30-Ti,
60MEA200/10iR-Ti, 60MEA200/30iR-Ti
Technical Specifications Standard 60MEA
Temperature
compatibility
0
-
125
°C
Dimension (W x D x H)
49 mm x 49 mm x 1 mm
Base
material
Glass
Track material
ITO (Indium tin oxide) or Ti (Titanium)
Contact pads
ITO (Indium tin oxide) or TiN (Titanium nitride)
Electrode diameter 10
or
30
μm
Interelectrode distance (centre to centre)
100 or 200 μm
Electrode
height
Planar
Electrode
type
TiN
(Titanium
nitride)
Isolation type
Silicon nitride 500 nm (PEVCD)
Electrode impedance
30 - 50 k for 30 μm electrodes, 250 - 400 k for 10 μm electrodes
Electrode layout grid
8 x 8
Number of recording electrodes
59 (with iR) or 60 (without iR)
Number of reference electrodes
1 internal reference electrode (iR) or without internal reference
MC_Rack
Source layout in “Data Source Setup”
2 dim. (MEA) or Configuration
Channel
map
Default
MEA perfusion chamber
(w/o)
Without ring
(gr)
Glass ring: ID +/- 19 mm, OD 24 mm, height 6 / 12 mm
(pr)
Plastic ring without thread: ID 26.5 mm, OD 30 mm, height 6 / 3 mm
(pr-T)
Plastic ring with thread: ID 26 mm, OD 30 mm, height 6 / 15 mm
21
31
41
51
61
71
12
13
14
15
16
17
22
23
24
25
26
27
28
32
33
34
35
36
37
38
42
43
44
45
46
47
48
52
53
54
55
56
57
58
62
63
64
65
66
67
68
72
73
74
75
76
77
78
82
83
84
85
86
87
49.0 mm
5.4 mm
200 / 100 μm
10 / 30 μm
2.2 mm
0.2 mm
Contact pads
Standard electrode layout grid 8 x 8
49.0 mm
REF