Layout with 12 x 12 electrode grid – Multichannel Systems MEA Manual User Manual
Page 120
![background image](/manuals/338871/120/background.png)
Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11
[email protected]
www.multichannelsystems.com
© 2013 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.
120pMEA200/30iR-Ti
120pMEA for Use with MEA2100-120-System
Layout with 12 x 12 electrode grid
GND
GND
GND
GND
46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75
45
44
43
42
41
40
39
38
37
36
35
34
33
32
31
30
29
28
27
26
25
24
23
22
21
20
19
18
17
16
120 119 118 117 116 115 114 113 112 111 110 109 108 107 106
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
REF
REF
REF
REF
C2 D3 D1 E3 E1 F3 F1 G5 G2 G4 H2 H4 J2 J4
G6 G1 G3 H1 H3 J1 J3 K2
C3 E5 D2 E4 E2 F4 F2 F5
K3
H5
L4
J5
L5
J6
L6
H6
G7
M7
K7
M8
K8
M9
K9
L10
L3
K4
M4
K5
M5
K6
M6
H7
L7
J7
L8
J8
L9
J9
B3
C4
A4
C5
A5
C6
A6
F6
D4
B4
D5
B5
D6
B6
E6
E7
B7
D7
B8
D8
B9
E8
C10
A7
C7
A8
C8
A9
C9
B10
D9 D11 E9 E11 F9 F11 F8
G12 G10 H12 H10 J12 J10 K11
G8 G11 G9 H11 H9 J11 H8 K10
C11 D10 D12 E10 E12 F10 F12 F7
Technical Specifications 120pMEA200/30iR-Ti
Temperature
compatibility
0
-
50
°C
Dimension (W x D x H)
49 x 49 x 1 mm
Base
material
Polyimide
foil
(2611)
with
perforation
on
glass
carrier
Perforation:
Diameter of innermost area
3 - 4 mm
Total area of holes
9-12 % (according to 3 - 4 mm)
Diameter of the holes
15, 20, 25, 30, 40, 45, 50 μm
Contact
pads
TiAu
(Titan,
Gold)
Track
material
TiAuTi
(Titan,
Gold,
Titan)
Electrode
diameter 30
μm
Interelectrode distance (centre to centre)
200 μm
Electrode
height
Planar
Electrode
type
TiN
(Titanium
nitride)
Isolation
type
Polyimide
foil
(2610)
isolator
Electrode
impedance
30
-
50
k
electrodes
Electrode
layout
grid
12
x
12
Number
of
recording
electrodes
120
Number
of
reference
electrodes
4
internal
reference
electrodes
(iR)
Contact pads for reference electrodes, connected to ground
4
MC_Rack
Source layout in “Data Source Setup”
Configuration
MEA
MEA120
200/30
Rinse with distilled water. Do not use ultrasonic bath!
These
perforated
MEAs
are
not
heat
stable,
and
should
not
be
autoclaved!
MEA perfusion chamber
(w/o) Without ring
(gr) Glass ring: ID +/-19 mm, OD 24 mm
(pr) Plastic ring without thread: ID 26.5 mm, OD 30 mm, height 3 / 6 mm